Menu

HIAC HRLD


Model : HIAC HRLD | Category : Lab Liquid Particle Counters | Brand : Beckman Coulter
The HIAC liquid particle counting HRLD series of light obscuration sensors are the industry leaders in reliability with a wide range of flow rates, high concentration limits and wide dynamic range. HIAC HRLD sensors can be used in both on-line and batch operation, and are highly adaptable for nearly any application.

HRLD series 100, 150, 400:
  • Hydraulic fluids
  • Phosphate esters
  • Water for injection
  • Common solvents
  • Kerosene
  • Diesel fuel
  • Jet fuel (JP4, JP5, and many others)
  • Purified water
  • Alcohols
HRLD series 150JA and 600JS are commonly used with:
  • Aldehydes
  • Ketones
  • Esters
  • Aromatics
  • Sodium hydroxide
  • Hydrogen peroxide
  • Ammonium hydroxide
*Viton™ seals standard. Kalrez™ seals available for phosphate esters.


HRLD Series Model Number Concentration Limit <10% coincidence) Flow Rates (mL/min)
HRLD-100 4 μm(c)-100 μm(c)* 10,000 20-100
HRLD-100HC 4 μm(c)-100 μm(c)* 18,000 10-50
HRLD-150 1.3 min.-150 18,000 10-25
HRLD-150JA 1.3 min.-150 18,000 10-25
HRLD-400 2.0 min.-400 10,000 20-100
HRLD-400HC 2.0 min.-400 18,000 10-50
HRLD-600JS 2.0 min.-600 6,000 30-200
* As per ISO 11171 and ISO 4406, note that 4 μm(c) < 2 μm
 
Pressure Limit 69 bar (1000 psi)
Sample Temperature Limit 65°C (150°F)
Calibration Options ASTM F658-87 (PSL spheres); ISO 4402 (ACFTD in oil); or ISO 11171 (MTD in oil). For pharmaceutical applications, sensor resolution can be factory tested in accordance with USP <788>.
When ordering, specify HRLD-100 or HRLD-100HC HRLD-150 or HRLD-150JA HRLD-400 or HRLD-400HC HRLD-600JS
Compliance Certifications CE Compliant per EMC and Low Voltage Directives Class 1 Laser Product Complies with IEC/EN 60825-1 and 21 CFR 1040.10 pursuant to Laser Notice 50