Menu

Met One A3432


Model : Met One A3432 | Category : Portable Particle Counters | Brand : Beckman Coulter
 
The MET ONE A3432 manifold system is designed for operation with the MET ONE 3400 series portable air particle counters. The MET ONE A3432 air particle counter manifold system allows a single particle counter to sample up to 32 locations.

Industry leading technology with the MET ONE 3400 series portable air particle counters
ISO 21501-4 reliability and accuracy Test Wizard for Easier Compliance, organize your clean room Flexible Communications – easy set up and data transfer Seamless Connectivity to Data Management Software. Economical clean room solution for facility monitoring
Sample up to 32 locations per manifold ISO class 1 to 8. Eliminate false positives due to sample line contamination
MET ONE A3432 air particle counting manifold system uniquely offers high flow purge through all sample lines. Random location sequencing
Sample high risk particle counter clean room locations more often.

Included

Part Number Description
2080359-01 Manifold
2083034-01 Manifold controller
400119 Controller cable
400100 AC power adapter 115 VAC (depending on location)
510667 AC power adapter 230 VAC (depending on location)
2083006-01 Manifold cable
Variations available Vacuum pump
2088927 Manifold cable adapter
 

Optional

Part Number Description
960380 Tubing
960011 Tubing, manifold to particle counter, ½" (optional)
204684-01 Coupling shoe
2082645-1 Isokinetic probe (optional)
2082644-3 Wall mount bracket for isokinetic probe (optional)

Spec Value
Environment  
Operating temperature 12 to 41 ºC (55 to 105 ºF)
Operating humidity 10% to 85% relative humidity, non-condensing
Storage temperature -40 to 71 ºC (40 to 160 ºF)
Storage humidity Up to 98% relative humidity, non-condensing
Maximum altitude 2000 meters (6560 ft)
Operating location Indoor use only
Manifold  
Number of ports 32
Sample tubing ½" OD, ³/8" ID
Size (W x H x D) 305 x 318 x 406 mm (12" x 12.5" x 16")
Weight 12.2 kg (26.9 lb)
Manifold controller  
Power 100-120 VAC, 750 mA max (T 1.0 A fuses) or 220-240 VAC, 375 mA max (T 0.5 A fuses) (specify voltage when ordering), 50/60 Hz
Size (W x H x D) 231 x 122 x 206 mm (9.1 x 4.8 x 8.1 in.)  
Weight 3.1 kg (6.7 lb)
Program cycle Up to 100 steps
Particle counter (additional)  
Data storage 400 samples, rotating buffer
Hold time 15 seconds minimum when manifold is attached
209E calculations On stored data
3-digit display Displays the port number when the location light is on
Compliance  
Certification UL, CSA and IEC certified by ETL, CE. Contact Hach for complete compliance details.
Copyright PT. Hartech Indonesia 2018