Menu

EZ-Fluo


Model : EZ-Fluo™ | Category : Bioburden Testing | Brand : Merck Millipore

EZ-Fluo™

Fast, non-destructive system for microbial detection

Overview

EZ-Fluo™ is a non-destructive method that enables you to continue to grow the microorganisms after they have been stained in order to identify them using any standard ID technology. This is a tremendous advantage compared to other rapid test systems which cannot ID microorganisms using the same sample that was tested to detect the contamination.
  • Rapid technology capable of providing results in 1/3 of the time required using traditional techniques
  • Flexible approach based on standard membrane filtration and reduced incubation on media plate
  • Non-destructive method, compatible with any ID technique
  • Compact hardware, fits easily on any laboratory bench
  • Cost effective solution for routine use as well as investigation

Faster product release

Today, more and more QC laboratories are switching to rapid methods for the detection of microbial contamination. This is because traditional methods require several days to provide results and therefore are of limited value when it comes to releasing product to the market faster, or to conducting quick investigation when necessary.

Applications

EZ-Fluo™ can detect and quantify microbial contamination in any filterable sample. Potential applications are numerous and include testing of the following sample types:
  • Raw materials
  • In-process samples
  • Final product
  • Environmental samples

Support and Services

We provide a comprehensive range of support and services to help you save your QC resources and facilitate the implementation of your EZ-Fluo™ system into your daily testing routine. Our highly experienced service teams can assist you with the following activities:
  • Feasibility study and on-site evaluation
  • Method development and method development
  • consulting
  • Equipment installation
  • Operator training
  • On-site system qualification (IQ/OQ)
  • Yearly preventative maintenance
  • Service plan
For further information on our support and services offering please contact your local sales representative.

Support and Services

Merck Millipore provides technical support and a comprehensive services offering to help you save your QC resources and facilitate the implementation of your EZ-Fluo™ into your daily testing routine. Our highly experimented teams are there to assist you.

For further information, please visit:
You can also contact your local sales representative.

Specifications

Dimensions (nominal) Component Length Width Height
  Reader 24.9 cm (9.8 in.) 14.2 cm (5.6 in.) 12.5 cm (4.9 in.)
  Camera 9.7 cm (3.8 in.) 6.6 cm (2.6 in.) 11.4 cm (4.5 in.)
Weight (nominal) Component Weight
  Reader 4.4 kg (9.7 lb.)
  Camera 0.5 kg (1.1 lb.)
Voltage Component Supply Voltage Input Supply Voltage Output
  Reader 100–240 V / 50–60 Hz 24V direct current
  Camera Powered by the computer
Power Component Consumption Power
    Nominal Maximal Nominal Maximal
  Reader 670 mA 2080 mA 16 watts 50 watts
  Camera 200 mA 430 mA 1 watt 2.15 watts
  Note: The reader is supplied with its power cord according to region.
Operating Temperature Range 15 to 40 °C
Humidity < 90 %
Altitude < 3,000 m (9,842 ft.)
Regulatory Information We certify that the EZ-Fluo™ Reader and the EZ-Fluo™ Camera were designed and manufactured in application of the following European Council directives:
  • Electromagnetic compatibility 2004/108/EC
  • Restriction of the use of certain Hazardous Substances in electrical equipment (RoHS) 2002/95/EC
Materials of Construction
Component Materials of Construction
Reader
Housing Painted aluminum sheet 1050
Optical chamber 304L stainless steel
Handle 304L stainless steel
Optical filter Glass
LED protector Transparent polycarbonate
Foot Polyvinylchloride (PVC)
Labels Polyester
Screen Protector polyester
Camera
Front plate 304L stainless steel
Housing Painted aluminium alloy
Window Glass
Light Guard Clear polymethyl methacrylate (PMMA)
Copyright PT. Hartech Indonesia 2018